Sarasota Industrial Density Meter FD910s
Specification
Transducer:
 
Operating range: 0 to 2.1g/cm3
Factory calibration range: 0.65 to 1.6g/cm3
Accuracy: ±0.0001g/cm3
Repeatability: 0.00002g/cm3
Temperature effect:
(corrected) see Note 1
0.000005g/cm3/°C
Process temperature: -50 to +180°C
Pressure effect:
(corrected) see Note 1
0.000003g/cm3/bar
Operating pressure: Flange pressure/temperature rating
Materials of construction: Sensor/wetted material – 316L stainless steel Case – 316L stainless steel Electronics housing – copper free aluminium grey epoxy finish
Secondary containment: Full pipeline specification at 40 bar
NACE MR0175 conformity: NACE specification option
Temperature measurement: High accuracy 1/3 DIN Integral 4 wire PT100
General:  
Environmental protection: IP65
Hazardous area certification: FD910F EEx ia IIC T6
FD910H EEx ia IIC T4
EEx d IIC T4/T3
Weight: Nett 11kg max
Shipping 20kg max
Installation: Vertical as standard
Optionally horizontal
Flow range: 0-300 l/min
5-300 l/min when installed horizontally
Outputs:  
F option: Frequency related to density on 2 wire current modulated loop 6mA to 18mA 4 wire PT100
H option: Analogue 4-20mA related to density or
density derived variable HART® protocol
System: Accuracy better than ±0.2% reading plus 0.05% of span at reference conditions HART® Local display option
Electrical:  
Power supply:  
F option: 13-28 VDC 10mA average (peak 18mA)
H option: 2 x 13-28 VDC 25mA
Connections: Screw terminals
Cable entry: 2 x 34h NPT
Operating temperature (ambient): -20 to +70°C
   
   
Ordering Information


Copyright 2002 Eastern Energy Services Pte Ltd